Input Calibration

Calibrating the Integral Pressure Sensor

Introduction

The concentration of oxygen dissolved in air-saturated water depends on the barometric pressure. This dependence is automatically compensated for during air calibration using a pressure sensor built into the Analyzer. The purpose of the pressure calibration is to calibrate that pressure sensor. However, this sensor has been factory calibrated and should not require re-calibration.

Procedure

Determine the true ambient barometric pressure, such as from a calibrated pressure transmitter or a mercury barometer. Absolute barometric pressure is required - not the “relative” sea-level pressure normally reported by the weather bureau.

Table 8-13 Calibrating the Integral Pressure Sensor

Step

 

Action

 

Screen

1

 

 

 

 

 

 

 

Press

Calibrate

 

CALIBRATION

 

 

 

Input PV Cal

 

 

 

 

 

 

Input Temp Cal

 

 

 

 

 

Output Cal

 

 

 

 

 

Cal History

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

 

 

 

 

 

 

 

Press

Enter

 

PV INPUT CAL

 

 

 

 

 

 

 

 

 

In 1 DO Cal

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Use

to select

 

 

 

Input 1 or 2 DO Cal

 

 

 

 

 

 

3

 

 

 

 

 

Enter

IN1 DO CAL

 

Press

Air Cal

 

 

 

 

Sample Cal

 

 

 

 

Reset Cal Factor

 

 

 

Pressure Cal

 

 

 

 

 

Pressure Offset

 

 

 

 

Reset Prs Offset

 

 

 

Bias Scan

 

 

 

 

Bias Volts

 

 

 

 

Reset Bias Volts

 

 

 

 

 

 

 

Use to select

Pressure Cal

January 2009

UDA2182 Universal Dual Analyzer Product Manual