Outputs Calibration

Procedure

Table 9-1 Procedure for Calibrating Analyzer Outputs

Step

Action

Screen

 

 

 

1Turn off the power to the Analyzer. More than one switch may be required to disconnect power.

2With the power off, open the case:

Loosen the four captive screws on the front of the bezel.

Grasp the bezel on the right side. Lift the bezel gently and swing the bezel open to the left.

3Refer to Figure 7-1 for the location of the terminal board retainer. Loose the screws that hold the retainer and slide the retainer right or left until the retainer tabs disengage from the terminal boards.

4Insert a screwdriver into the tab in the terminal board to be wired and pull. Slide the board half way out. - Refer to Figure 7-1 for the location.

(Output 1 and 2 – Power Supply/Analog Output/Relay Output card)

(Output 3 – Option card)

There is a notch in the terminal board into which you can slide the retainer tabs and hold the boards in place while wiring.

5Label and remove the field wiring from the output terminals. Output 1 – Terminals 12– and 13+

Output 2 – Terminals 10– and 11+ Output 3 – Terminals 14– and 15+

6Feeding the test wiring through the conduit hole in the case and connect the appropriate type meter to the specific output terminals Be sure to observe the correct polarity.

Slide the Input board back and close the case and power up the unit. Do not apply power until the case is closed.

7

Calibrate

CALIBRATION

Press

 

 

Input PV Cal

 

 

 

 

Input Temp Cal

 

 

Output Cal

 

 

 

Cal History

 

 

 

 

 

 

 

 

 

Use to select

”Output Cal”