Configuration

Sub-menu

Parameter

Selection or

 

Parameter Definition

selection

 

Range of

 

 

 

 

 

 

Setting

 

 

 

 

 

Solution pH/ºC

0.0 (default) to

Measured pH is displayed and transmitted normalized

 

(Temp Input =

-0.050 pH/ºC

to a solution temperature of 25°C as determined by

 

 

the current Solution Temperature Coefficient. This is

 

Enable, Solu

 

 

 

expressed in units of pH/°C with precision to the

 

Temp Comp =

 

 

 

hundredths decimal place. The parameter “Solu Temp

 

Custom)

 

 

 

Coeff” allows the selection of the following entries.

 

 

 

 

 

 

Follow the “General Rules for Editing” in section

 

 

 

6.4.1 to make the changes. (-) Will appear when first

 

 

 

digit to the right of decimal point is changed.

 

 

 

 

Solution Type

Temp Coefficient

 

 

 

 

 

 

 

 

 

 

 

None (Default)

0.000

 

 

 

 

 

H2O (Pure Water)

-0.016

 

 

 

 

 

NH3 (Ammonia)

-0.032

 

 

 

 

 

PO4 (Phosphate)

-0.032

 

 

 

 

 

C4H9NO (Morpholine)

-0.032

 

 

 

 

 

Custom

User Entry

 

 

PV Bias

-99999.00 to

PV Bias Constant - is used to compensate the input

 

 

99999.00

for drift of an input value.

 

 

 

 

default = 0.00

 

 

 

 

 

 

 

 

 

Failsafe

-99999.00 to

The output value to which the output will go to protect

 

 

99999.00

against the effects of failure of the equipment.

 

 

default = 14.00

 

 

 

 

 

 

 

 

 

Filter Time

0 to 120

A software digital filter is provided for dampening the

 

 

default = 0.0

process noise. This filter is applied before the limit

 

 

 

functions.

 

 

 

 

 

 

 

 

 

66

UDA2182 Universal Dual Analyzer Product Manual

January 2009

Page 78
Image 78
Honeywell UDA2182 manual Default to