Diagnostics and Messages

12.3 Calibration Diagnostics

pH/ORP/DO

All of the possible errors are detected during a probe calibration and will abort the calibration process with the message “FAIL” appearing briefly, followed by a return to the online pH/ORP/DO display. At that point, the specific error will be displayed as described. In addition, any of following errors may occur during probe calibration and abort the calibration process.

Table 12-2 Probe Calibration Diagnostics

Status Message

Definition

BUFFER SPAN TOO LOW

The span between pH buffer 1 and pH buffer 2 is less than 2 pH. Use a set of

 

buffers that are at least 2 pH apart. As a warning status, will clear when an

 

appropriate buffer 2 value is selected. As an error message, will abort

 

calibration and preserve original slope value.

 

 

OFFSET UNDERRANGE

Resulting pH offset (standardization) value is less than –2 pH after pH slope

 

calibration. Calibration is aborted and original pH offset and slope values are

 

preserved.

 

 

OFFSET OVERRANGE

Resulting pH offset (standardization) value is greater than 2 pH after pH slope

 

calibration. Calibration is aborted and original pH offset and slope values are

 

preserved.

 

 

SLOPE UNDERRANGE

Resulting pH slope is less than 80%. Calibration is aborted and original slope

 

value is preserved.

 

 

SLOPE OVERRANGE

Resulting pH slope is greater than 105%. Calibration is aborted and original

 

slope value is preserved.

 

 

CAL FACTOR UNDERRANGE

Resulting DO calibration factor is less than 0.001268. DO calibration is

 

aborted and original calibration factor is preserved.

 

 

CAL FACTOR OVERRANGE

Resulting DO calibration factor is greater than 0.040580. DO calibration is

 

aborted and original calibration factor is preserved.

 

 

PROBE CURRENT TOO LOW

DO probe current is less than 5 µA. DO bias scan is aborted and original bias

 

voltage is preserved.

 

 

PROBE CURRENT TOO HIGH

DO probe current exceeds the greater of 133% of the probe current at last

 

successful calibration or 160 µA. During DO bias scan, scan is aborted and

 

original bias voltage is preserved.

 

 

READINGS UNSTABLE

DO air PV or temperature readings too unstable for successful air calibration.

 

Calibration is aborted and original calibration factor is preserved.

 

 

SOLUTION UNSTABLE

pH solution PV or temperature readings too unstable for successful auto

 

buffer calibration. Calibration is aborted and original pH offset (for buffer 1) or

 

slope value (for buffer 2) is preserved.

 

 

SOLUTION TEMP TOO LOW

pH solution temperature readings less than minimum of 0 degrees C. Auto

 

buffer calibration is aborted and original pH offset (for buffer 1) or slope value

 

(for buffer 2) is preserved

 

 

SOLUTION TEMP TOO HIGH

pH solution temperature readings greater than maximum of 100 degrees C.

 

Auto buffer calibration is aborted and original pH offset (for buffer 1) or slope

 

value (for buffer 2) is preserved

 

 

January 2009