Input Calibration

Procedure

Make sure you have selected “PV Type –pH Glass, pH Durafet, or pH HPW” in the Inputs configuration -Table 6-5.

Refer to Section 6.4.1 – General Rules for Editing.

Table 8-2 Calibrating pH Electrodes Using Automatic Buffer Recognition

Step

Action

Screen

 

 

 

1Prepare containers of two standard reference solutions.

2

Calibrate

 

CALIBRATION

 

 

 

 

 

Press

 

Input PV Cal

 

 

 

 

 

 

 

Input Temp Cal

 

 

 

 

 

 

Output Cal

 

 

 

 

 

 

Cal History

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Use

to select

 

 

 

 

 

Input PV Cal

 

 

 

 

 

 

 

 

3

 

 

 

 

 

 

 

Enter

 

PV INPUT CAL

 

 

Press

 

 

 

 

 

 

In 1 pH/ORP Cal

 

 

 

 

 

 

 

 

 

 

In 2 Conduc Cal

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Use

to select

 

 

 

Input 1 or 2 pH/ORP Cal

 

 

 

 

 

 

4

 

 

 

 

Enter

IN 1 pH/ORP Cal

 

Press

 

 

 

Auto Buffer Cal

 

 

 

 

Buffer Cal

 

 

 

 

Sample Cal

 

 

 

 

 

 

 

 

Buffer Group

 

 

 

 

pH Offset

 

 

 

 

 

 

 

 

pH Slope

 

 

 

 

Reset pH Offset

 

 

 

 

Reset pH Slope

 

 

 

 

 

 

 

 

 

Use

to select

 

 

”Buffer Group”

 

 

 

 

5

Enter

Use

to select

Press

NIST/USP (default)

 

 

USA, or

Europe

January 2009

UDA2182 Universal Dual Analyzer Product Manual

143