Input Calibration

8.7.5Performing Calibration Trim Introduction

For most applications entering the cal factor for each cell will achieve satisfactory system performance. However, it is possible to perform a calibration trim procedure in which the Analyzer and cell combination are used to measure a reference solution of known conductivity; the reading of the Analyzer is adjusted to match.

The same procedure can be used to adjust the Analyzer’s reading while the cell is in the process, if a reference instrument is used to determine the process conductivity. In this case, the process fluid becomes the “reference solution”.

Calibration trim is recommended for acid concentration measurements above 5%.

Materials

To perform calibration trim using a standard reference solution, follow the instructions in Table 8-9.

Materials required are:

A reference solution of known conductivity near the point of interest, with the temperature controlled (or measured and compensated) to within ± 1 °C. Conductivities of potassium chloride solutions are provided in Table 8-8. Solutions must be prepared with high-purity de-ionized, CO2-free water, and dried potassium chloride.

For acid concentration applications, a certified reagent grade solution with the temperature controlled.

A container for the reference solution, large enough to immerse the cell to measuring depth.

De-ionized water to rinse the cell.