Honeywell UDA2182 Modbus Communications, Summary, Serial port provides, Ethernet port provides

Models: UDA2182

1 258
Download 258 pages 46.92 Kb
Page 62
Image 62

Operating the Analyzer

5.13 Modbus Communications

Overview

The UDA2182 provides Modbus communication support on two communication interfaces using the optional Communication Card. A general summary is listed below.

For details see UDA2182 Communications User Guide #70-82-25-126.

Summary

Communications Card (Optional)

The Communications card provides one Serial Port (RS485) and one Ethernet Port.

Serial port provides

RS422/RS485 multi-drop

1200 to 38400 programmable baud rate

Modbus RTU protocol to read signals including PV, Temperature, Alarm Status, outputs, relay status, etc.

Read/write four analog and four digital variables (Note 1)

Ethernet port provides:

Up to 5 Modbus TCP connections simultaneously

Ethernet parameters are configured via the front-panel or web pages.

Web server with up to10 clients simultaneously

Multi-language Web pages (Note 2) setup the Ethernet port settings and monitor readings, alarms, statuses, events

Multi-language Email to send alarm status changes. Alarm notification to eight email addresses. These must be configured using web pages signed in as the administrator.

DHCP: ( Dynamic Host Configuration Protocol) selectable via web page or front-panel

Firmware upgrade to Main CPU board

Firmware upgrade to Communications card

Note 1

There are four analog and four digital variables. These variables can be read and written remotely using Modbus function codes.

Variables will appear as a selection for various parameters:

Analog variables can be an alarm source, analog relay source, current output source, monitor source, math source, auxiliary switch source, PID TRV, and PID remote setpoint.

Digital variables can be an alarm disable, remote setpoint select, Tune Set2 select, digital relay source, logic- in source, auxiliary switch select, PID TRC select, PID RSP select, and auto cycle start source

Note 2

Web pages provide the following:

Multiple language support

“Guest” accessibility for read-only permission

“Admin” accessibility for read and write permission

Readings of Inputs, Outputs, and Relay Outputs

Status of Inputs, Outputs, and Alarms.

Readings and Status of optional parameters (control, pharma, and auto-cycle)

List of last twelve events

Network configuration including IP address, subnet mask, gateway etc.

Email configuration for alarm event notification

50

UDA2182 Universal Dual Analyzer Product Manual

January 2009

Page 62
Image 62
Honeywell UDA2182 manual Modbus Communications, Summary, Serial port provides, Ethernet port provides