Honeywell UDA2182 manual NIST-default +, Cond mS/m, ISO-Default +

Models: UDA2182

1 258
Download 258 pages 46.92 Kb
Page 80
Image 80

Configuration

Sub-menu

Parameter

Selection or

Parameter Definition

selection

 

Range of

 

 

 

Setting

 

 

 

 

 

PV Type

Cond

μ

These selections are only available with regard to the

 

S/cm

 

 

 

Select Cell

(NIST-default)+

Cell Constant selected (See “Cell Constant”).

Cond mS/cm –

 

 

 

Constant First

Cell

Available Selectable PV Types

(NIST)

 

Constant

Use the

keys to select

 

Concentrtn

 

 

TDS ppb

0.01

Conductance

μ

 

TDS ppm

S/cm (default- NIST), Conductance

 

 

mS/cm (NIST), Conductance mS/m(default - ISO),

 

TDS ppt

 

Conductance S/m (ISO), TDS ppb, TDS ppm,

 

Resistivity

 

Resistivity, Conductance μS/m (ISO)

 

Cond mS/m

0.1

Conductance μS/cm (default- NIST), Conductance

 

(Default)

mS/cm (NIST), Conductance mS/m(default - ISO),

 

(ISO-Default)+

 

Conductance S/m (ISO), TDS ppb, TDS ppm,

 

Cond S/m –

 

 

 

Resistivity, Conductance μS/m (ISO)

 

(ISO)

 

 

 

 

 

1

 

Conductance μS/cm (default- NIST), Conductance

 

Concentrtn

 

 

 

 

 

mS/cm (NIST), Conductance mS/m(default - ISO),

 

TDS ppb

 

 

 

Conductance S/m (ISO), TDS ppm, TDS ppt,

 

TDS ppm

 

 

 

 

 

10

 

Conductance μS/cm (default- NIST), Conductance

 

TDS ppt

 

 

 

 

 

mS/cm (NIST), Conductance mS/m(default - ISO),

 

Resistivity

 

 

 

 

 

 

 

Conductance S/m (ISO), TDS ppm, TDS ppt.

 

Cond μS/m

 

 

 

 

 

25

 

Concentration (default), Conductance μS/cm

 

 

 

 

 

 

 

 

(default- NIST), Conductance mS/cm (NIST),

 

+ parameter

 

 

 

Conductance mS/m(default - ISO), Conductance

 

 

 

 

S/m (ISO),

 

selected in

 

 

 

 

 

50

 

Concentration (default), Conductance μS/cm

 

MAINTENANCE

 

 

 

 

 

(default- NIST), Conductance mS/cm (NIST),

 

Æ INPUTS menu

 

 

 

 

 

 

 

Conductance mS/m(default - ISO), Conductance

 

 

 

 

 

S/m (ISO),

 

 

 

 

 

 

 

 

 

 

PV Range

 

Read Only

 

 

 

 

Cell Constant *

0.01

The Cell Constant is a value specific to a category of

 

0.1 (default)

cells for the measurement range required.

 

1

 

 

 

 

 

10

 

 

 

 

 

25

 

 

 

 

 

50

 

 

 

 

 

 

 

Cal Factor *

0.850 to 1.150

The Cal Factor is a correction value applied to the

 

default = 1.000 cell’s Cell Constant, which is unique to each cell to

 

 

take into account tolerances in manufacture.

If a standard cell is attached to the sensor, the Cell Constant defaults to “0.1” and the Cal Factor defaults to “1.000”. These standard cell parameter values are editable and are retained through a power cycle.

*Cell Constant and Cal Factor are automatically uploaded from Honeywell conductivity cells with EEPROM (blue & brown leads) and these values cannot be edited.

68

UDA2182 Universal Dual Analyzer Product Manual

January 2009

Page 80