Configuration

 

 

 

 

 

 

Sub-menu

 

Parameter

Selection or Range

Parameter Definition

selection

 

 

of Setting

 

 

 

Address

0 to 999 (default = 0)

Modbus RTU Slave ID – 0 is offline

 

 

 

 

 

Baud Rate

2400 (default)

Modbus RTU Baud Rate

 

 

 

4800

 

 

 

 

 

9600

 

 

 

 

 

19200

 

 

 

 

 

38400

 

 

 

 

 

57600

 

 

 

 

 

115200

 

 

 

 

 

 

Ethernet

Port Reset

Off (default)

Enable selection resets the Communication card.

 

 

 

Enable

It should be enabled when configurations for

 

 

 

 

Ethernet are modified.

 

 

 

 

 

DHCP

No (default)

When YES, Dynamic Host Configuration Protocol

 

 

 

Yes

server automatically assigns a dynamic IP

 

 

 

 

address to UDA.

 

 

 

 

The set dynamic IP can be seen from “Comm

 

 

 

 

Status” display.

 

 

 

 

 

IpAddr Octet 1

0 to 255

These parameters are visible only when DHCP

 

IpAddr Octet 2

 

option is NO.

 

IpAddr Octet 3

 

Allows you to assign Static IP address to the

 

IpAddr Octet 4

 

 

 

UDA.

 

 

 

 

 

 

 

 

 

SbntMsk Octet 1

0 to 255

These parameters are visible only when DHCP

 

SbntMsk Octet 2

 

option is NO.

 

SbntMsk Octet 3

 

Allows you to assign Subnet Mask as per the

 

SbntMsk Octet 4

 

 

 

local network settings

 

 

 

 

 

 

 

 

 

Dflt Gtwy Octet 1

0 to 255

These parameters are visible only when DHCP

 

Dflt Gtwy Octet 2

 

option is NO.

 

Dflt Gtwy Octet 3

 

Allows you to assign the Default Gateway as per

 

Dflt Gtwy Octet 4

 

 

 

the local network settings

 

 

 

 

 

 

 

 

 

 

 

DNS Srvr Octet 1

0 to 255

These parameters are visible only when DHCP

 

 

DNS Srvr Octet 2

 

option is NO.

 

 

DNS Srvr Octet 3

 

Allows you to assign the DNS server IP address

 

 

DNS Srvr Octet 4

 

 

 

 

as per the local network settings

 

 

 

 

 

 

 

 

 

 

January 2009

UDA2182 Universal Dual Analyzer Product Manual

107