Operating the Analyzer

 

 

 

 

 

Status

Parameter

Status

Status Definition

Display

 

(Read Only)

 

 

Math Values

Math 1

Analog Values

Read Only – Shows the calculated values of the Math

 

Math 2

 

blocks.

 

Math 3

 

 

 

 

Math 4

 

 

 

 

 

 

 

Aux Values

Switch 1

Analog Values

Read Only – Shows the calculated values of the

 

Switch 2

 

blocks in the Aux Group. This includes the current

 

Func Gen 1

 

output of the Switch and Function Generator blocks.

 

Func Gen 2

 

 

 

 

 

 

 

Variables

AnlgVar 1

Analog Values

Read Only – shows values of Analog and Digital

 

AnlgVar 2

 

variables written from Modbus client.

 

AnlgVar 3

 

 

 

 

AnlgVar 4

 

 

 

 

Dgtl Var 1

Digital Values

 

 

 

Dgtl Var 2

(On or Off)

 

 

 

Dgtl Var 3

 

 

 

 

Dgtl Var 14

 

 

 

 

 

 

 

Comm Status

Comm Card Stat

Ok/

Read Only – This displays the status of the

 

 

Not Present/

Communication card. Information present only if the

 

 

HW Failure/

Communication card is present.

 

 

Fail

Comm Card Stat gives the status of the

 

 

Init

communication card

 

 

 

Status shown OK if the communication card is

 

 

 

working fine.

 

 

 

Status shown as Not present if the communication

 

 

 

card is not present.

 

 

 

Status shown as HW Failure if the communication

 

 

 

card is installed but unable to communicate to the

 

 

 

Main CPU board.

 

 

 

Status shown as Fail if the communication card is not

 

 

 

functioning properly. It could be the result of a

 

 

 

software failure, a bad flash chip on the board or

 

 

 

DHCP is selected but the DHCP server was not found.

 

 

 

Check cable connections and potential network issues

 

 

 

for DHCP related problems.

 

 

 

Status shown as INIT if the communication card is

 

 

 

getting initialized.

 

SW Version

Value

SW Version gives the software version of the

 

 

 

Communication Card.

 

 

 

 

 

Web Page

Value

Web Page gives the web page version number. The

 

 

 

web pages are separate from the Communication card

 

 

 

firmware, and can be upgraded independently. Both

 

 

 

the SW version and Web Page version should be the

 

 

 

same value to guarantee compatibility.

 

 

 

 

 

January 2009

UDA2182 Universal Dual Analyzer Product Manual

43

Page 55
Image 55
Honeywell UDA2182 manual Aux Values, Output of the Switch and Function Generator blocks, Variables, Comm Status