Input Calibration

Step

 

Action

 

Screen

6

 

 

 

 

 

 

Press

Enter

 

IN 1 pH/ORP Cal

 

 

 

Auto Buffer Cal

 

 

 

 

 

 

 

 

 

Buffer Cal

Sample Cal

Buffer Group

pH Offset

pH Slope

Reset pH Offset

Reset pH Slope

Use to select

”Auto Buffer Cal”

7

Put the unit in “Hold”

 

 

 

 

mode

 

 

 

Remove the electrode

 

 

 

 

from the process.

 

 

 

Rinse the electrode

 

 

 

 

thoroughly with distilled or

 

 

 

 

de-ionized water

 

 

 

 

 

8

Calibrating the pH Offset

“Place probe in Buffer 1”

 

Press Enter

The display will show the pH of

 

the buffer 1 solution as

 

Follow the prompts at the top

measured by the electrode

 

system.

 

 

and bottom of the screen.

 

 

The reading will be automatically

 

 

 

 

 

 

adjusted to match the known pH

 

 

 

value stored in the UDA2182

 

 

 

memory.

 

 

 

“Press Enter when stable”

 

 

 

9

Once the reading is stable

“Buffer 1 stability check”

 

Press Enter

Use

to change the value of

 

the Buffer.

 

 

 

“Up/Down changes Buffer”

10Rinse the electrode thoroughly with distilled or de-ionized water

11

Calibrating the Slope

“Place probe in Buffer 2”

 

Press Enter

The display will show the pH of

 

the buffer 2 solution as

 

Follow the prompts at the top

measured by the electrode

 

system.

 

and bottom of the screen.

 

The reading will be automatically

 

 

 

 

adjusted to match the known pH

 

 

value stored in the UDA2182

 

 

memory.

 

 

“Press Enter when stable”

144

UDA2182 Universal Dual Analyzer Product Manual

January 2009

Page 156
Image 156