Honeywell UDA2182 manual Resume Dly Mins

Models: UDA2182

1 258
Download 258 pages 46.92 Kb
Page 116
Image 116

Configuration

Sub-menu

Parameter

Selection or Range

Parameter Definition and Restrictions

selection

 

of Setting

 

 

Max Cal Mins

2

If the reading is unstable after 2 minutes then get a

 

 

 

“AUTOCYCLE FAIL ALARM”.

 

 

 

 

 

Resume Dly Mins

5

Wait 5 minutes after cycle completes and sensor is

 

 

 

reinserted before removing HOLD and returning to

 

 

 

On-Line mode.

 

 

 

 

104

UDA2182 Universal Dual Analyzer Product Manual

January 2009

Page 116