Configuration

Sub-menu

Parameter

Selection or Range

 

Parameter Definition and Restrictions

selection

 

of Setting

 

 

 

Start Day

1 to 28 (default = 1)

 

Cycle Interval is Monthly

 

(Dependent

Sunday – Saturday

 

Cycle Interval is Weekly

 

parameters)

(default = Sunday)

 

 

 

 

1 to 31 (default = 1)

 

Cycle Interval is Custom, Start Time enabled

 

 

 

 

 

 

Start Hour

0 to 23

 

Cycle Interval is Monthly, Weekly, Daily

 

(Dependent

(default = 12)

 

 

 

 

 

 

 

parameters)

 

 

 

 

 

 

 

 

 

Start Mins

0 to 59 (default = 0)

 

Cycle Interval is Monthly, Weekly or Daily

 

(Dependent

 

 

Cycle Interval is Custom, Start Time enabled

 

 

 

 

 

parameters)

 

 

 

 

 

 

 

 

 

Period Days

0 to 100(default =

 

The period day parameter allows the selection

 

(Custom)

0)

 

of how often the Auto cycle will occur.

 

 

 

For Example: 20 means that the Auto Cycle

 

 

 

 

 

 

 

 

will occur every 20 days.

 

 

 

 

 

 

Period Hours

0 to 23 (default = 1)

 

For Example: 4 means that the Auto-cycle will

 

(custom)

 

 

occur every 4 hours when the days and

 

 

 

minutes are set to 0.

 

 

 

 

 

 

 

 

 

 

Period Mins

0 to 59 (default = 0)

 

For Example: 30 means that the Auto-cycle

 

 

 

 

will occur every 30 minutes when the days

 

 

 

 

and hours are set to 0.

 

 

 

 

 

 

Rinse Cycle Cnt

0 to 100(default =

 

Allows selection for frequency of rinse occurrence.

 

 

1)

 

 

 

 

 

 

 

 

Cal Cycle Cnt

0 to 100(default =

 

Allows selection for frequency of calibration

 

(PV is DO)

1)

 

occurrence. For Example: 1 indicates that a

 

 

 

calibration will occur every cycle while a 10

 

 

 

 

indicates that a calibration will occur every 10th

 

 

 

 

cycle.

 

 

 

 

 

 

Cal 1 Cycle Cnt

0 to 100(default =

 

Allows selection for frequency of calibration

 

(PV is pH)

1)

 

occurrences.

 

 

 

 

 

 

 

 

 

 

Cal 2 Cycle Cnt

0 to 100(default =

 

Allows selection for frequency of calibration

 

(PV is pH)

1)

 

occurrences.

 

 

 

 

 

 

 

 

 

 

Max Transit Mins

0 to 30.00

 

Maximum Probe Transit time in minutes.

 

(Probe Transit =

(default = 0.50)

 

 

 

 

 

 

 

Enabled)

 

 

 

 

 

 

 

 

 

Rinse Mins

0 to 30.00

 

Duration of Rinse sequence in minutes

 

 

(default = 0.50)

 

 

 

 

 

 

 

 

Max Cal Mins

0 to 30.00

 

Maximum calibration time in minutes. Enough time

 

(PV pH, DO)

(default = 0.50)

 

should be entered to allow process stabilization

 

 

 

and a 20 second measurement time.

102

UDA2182 Universal Dual Analyzer Product Manual

January 2009

Page 114