Temperature Input Calibration

10.2 Temperature Input Calibration

Introduction

Temperature Input Calibration lets you monitor a live temperature reading while continuing to monitor the sample. The currently displayed temperature value can be edited through a series of prompts on the screen. The temperature offset value is always displayed in the temperature units selected in the Maintenance setup menu.

Accessing the Main Calibration Menu and sub-menus

Press Calibrate . The Main Calibration Menu will appear.

CALIBRATION

Input PV Cal

Input Temp Cal

Output Cal

Cal History

Use the keys to highlight the “Input Temp Cal” selection.

Press Enter to display the sub-menu for that selection.

Procedure

Table 10-1 Procedure for Calibrating the Temperature Inputs

Step

 

Action

 

Screen

 

 

 

 

 

 

 

1

 

 

 

 

 

 

Press

Calibrate

 

CALIBRATION

 

 

 

Input PV Cal

 

 

 

 

 

Input Temp Cal

 

 

 

 

 

 

Output Cal

 

 

 

 

 

 

Cal History

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Use

to select

 

 

 

 

 

” Input Temp Cal”

 

 

 

 

 

 

 

 

 

2

 

 

 

 

 

 

Enter

TEMP INPUT CAL

 

 

 

Press

 

 

 

 

 

T1 pH/ORP Cal

 

 

 

 

 

 

T2 Conduc Cal

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

186

UDA2182 Universal Dual Analyzer Product Manual

January 2009

Page 198