Input Calibration

 

 

 

 

 

 

 

Step

Action

 

Screen

 

 

6

Once the reading is stable

“Change to Buffer 1 value”

 

 

 

Press Enter

Use

to change the value to

 

 

match the actual pH of the

 

 

 

Buffer 1 solution at its current

 

 

 

temperature.

 

 

 

“Enter to save, Exit to cancel”

7Rinse the electrode thoroughly with distilled or de-ionized water.

8

Percent Theoretical Slope

“Place probe in Buffer 2”

 

Adjustment

The display will show the pH of

 

Press Enter

the buffer 2 solution as

 

measured by the electrode

 

Follow the prompts at the top

system.

 

 

“Press Enter when stable”

 

and bottom of the screen.

 

 

 

 

 

 

9

Once the reading is stable

“Change to Buffer 2 value”

 

Press Enter

Use

to change the value to

 

match the actual pH of the Buffer

 

 

2 solution at its current

 

 

temperature.

 

 

“Enter to save, Exit to cancel”

 

 

 

10

If the calibration fails, an error

Error Messages:

 

message will be displayed

Buffer span too low

 

across the bottom stripe of the

 

 

 

 

screen.

Slope (Percent Theoretical

 

Make necessary adjustments

Slope) underrange

 

 

 

 

and re-calibrate

Slope(Percent Theoretical

 

 

Slope) overrange

 

 

Solution Unstable

 

 

See Table 12-2 for definitions

January 2009

UDA2182 Universal Dual Analyzer Product Manual