6.Installation
6.1Site Requirements
The place of installation must meet the following requirements:
•No direct sunlight.
•Power supply at a distance no greater than approx. 3 m.
•No drafts (air condition outlets etc.).
•Even floor.
•
•The instrument is designed for indoor use only.
•
•The power switch/circuit breaker must be freely and easily accessible.
•Room temperature always approx. 22°C.
•The relative humidity must not exceed 60%.
To ensure proper function of the instrument, it must be set up while maintaining a minimum distance from walls and furniture.
• Distance to walls and furniture, calculated from the cabinet:
- rear: | 15 cm |
- right side: | 30 cm |
- left side: | 15 cm |
• No heat dissipating appliances around.
High room temperatures and excessive air humidity affect the cooling capacity of the cryostat and lead to ice forming in the instrument!
• Plug the instrument into power sockets with ground only. Power cable length: up to 3.5 m
Extendable: no
20 | Operating Manual V1.1 – 05/2007 |