8. Working with the Instrument

8.4.2 Trimming with Extraction – Anti-roll Guide Installed

E

Fig. 38

Remove the silicone stopper from the filter cover (and keep it in a safe place).

Insert the extraction hose with the black adapter.

Fold the anti-roll guide to the side and fasten the suction nozzle to the pressure plate – see mark Fig. 39a - (using 4 magnets on the rear side of the nozzle).

Fold the anti-roll guide back into position.

Fig. 39a

K B D

Fig. 39b

Fig. 39c

Ensure that the hose with the nozzle is not installed against its "natural" cur- vature on the pressure plate of the blade holder.

C

I

A

Fig. 40

The tension acting on the hose can be minimized by turning the red ring (Fig. 39, top right) clock- wise so that the suction nozzle presses against the pressure plate (I, Fig. 39).

Check that the anti-roll guide is parallel and correctly adjusted. Readjust if necessary (see page 58).

Activate trimming mode.

Select the trimming thickness.

Switch on the VAC and select a low extraction level (between 1 and 2).

Start trimming by manually moving the handwheel or press the RUN/STOP and ENABLE keys simultaneously to start motorized sectioning.

Leica CM1950 – Cryostat

43

Page 43
Image 43
Leica CM1950 manual Trimming with Extraction Anti-roll Guide Installed