| 3. | Technical Data |
|
|
| |
Refrigeration system | CM1950, 50 Hz | CM1950, 60 Hz | |
Cryochamber |
|
|
|
Temperature range | 0°C to | 0°C to | |
| at ambient temperature of 20°C | ambient temperature of 20°C | |
Cooling time to | approx. 5 h | approx. 5 h |
|
Cooling time to | approx. 8 h | approx. 8 h |
|
Refrigeration capacity ① | 690 W | 690 W |
|
25 bar | 25 bar |
| |
Safety factor | 3 | 3 |
|
Refrigerant* | 300 g (± 5 g) refrigerant R | 300 g (± 5 g) refrigerant R | |
Compressor oil* | 0.6 l EMKARATE RL22S, ICI* | 0.6 l EMKARATE RL22S, ICI* | |
Defrosting of cryochamber |
|
|
|
Automatic defrosting |
|
|
|
Programmable: | Yes (hot gas defrost) Selectable time | Yes (hot gas defrost) Selectable time | |
Defrosting intervals | 1 defrost in 24 h | 1 defrost in 24 h |
|
| or manual hot gas defrost | or manual hot gas defrost | |
Defrost time: | 12 minutes | 12 minutes |
|
Automatic shutoff Defrost: | at | at | |
|
|
| |
Minimum temperature: | - 42 °C (+ 5 K), at chamber temp. | - 42 °C (+ 5 K), at chamber temp. | |
Number of freezing stations: | 15+2 | 15+2 |
|
Defrost | Manual | Manual | |
Peltier element |
|
|
|
Number of freezing stations: | 2 | 2 |
|
Max. temperature difference: | 17 K, at chamber temp. of | 17 K, at chamber temp. of | |
Specimen cooling |
|
|
|
Temperature range |
| ||
Refrigeration capacity① | 320 W | 320 W |
|
25 bar | 25 bar |
| |
Safety factor | 3 | 3 |
|
Refrigerant and quantity | at 230V/50 Hz 140 g (± 5 g), refrigerant | at 120V/60 Hz 110 g (± 5 g), refrigerant | |
| at 100V/50/60 Hz 110 g (± 5 g), refrigerant |
|
|
Compressor oil* | 0.4 l alpha 22, Kyodo* | 0.4 l alpha 22, Kyodo* | |
Defrosting of specimen head |
|
|
|
Automatic defrost: | No | No |
|
Manual defrost |
|
|
|
Defrost time | 15 min. | 15 min. |
|
*) Refrigerant and compressor oil must be replaced by qualified, authorized service personnel only.
①acc. to CECOMAF: liquid temperature 45 °C, evaporation temperature:
Leica CM1950 – Cryostat | 15 |