12. Maintenance and Cleaning
Basically, we recommend UV disinfection (see page 35 for detailed instructions).
For
The cryostat has to be disinfected after each daily use.
Comply with the instructions for use!
The glass
It is not necessary to oil the parts, such as the
•In case of visible pollution (such as dust), clean the air inlet opening (Fig. 10, p. 18) of the condenser on the bottom
Do not turn the instrument on before the cryochamber is completely dry!
Frost formation!
The front panel and the slit cover of the microtome must be com- pletely dry before turning on the instrument!
Dry all parts completely before reinserting them in the cryochamber!
12.2 Changing Fuses
•In case of power supply faults, please contact an authorized Leica service technician immediately.
Do not carry out any repairs on your own as this will invalidate the warranty.
Repairs may only be carried out by qualified service engineers authorized by Leica.
Leica CM1950 Cryostat | 67 |