9.Troubleshooting

9.1Problems During Work

Problem

Causes

Remedies

 

 

 

Frost on chamber walls and microtome

Cryostat is exposed to air currents (open windows and doors, air conditioning).

Frost built-up by breathing into the cryochamber.

Change place of installation for the cryostat.

Wear mouth protection.

Sections smear

Specimen not cold enough.

 

Anti-roll plate not yet cold enough,

 

thus warming the section.

Select lower temperature. Wait until knife and/or anti-roll plate have reached chamber temperature.

Sections splinter

Specimen too cold.

Select higher temperature.

 

 

 

Sections not properly

Static electricity/air currents.

Remove cause.

flattened

Specimen not cold enough.

Select lower temperature.

 

Large area specimen.

Trim the specimen parallel,

 

 

increase section thickness.

 

Anti-roll plate poorly positioned.

Reposition anti-roll plate.

 

Anti-roll plate poorly aligned with knife

Align correctly.

 

edge.

 

 

Incorrect clearance angle.

Set correct angle.

 

Blunt knife.

Use different part of the knife.

 

 

 

Sections not properly flattened despite correct temperature and correctly aligned anti-roll plate

Dirt on knife and/or anti-roll plate. Edge of anti-roll plate damaged. Blunt knife.

Clean with dry cloth or brush. Replace anti-roll plate.

Use different part of the knife.

Sections curl on anti-roll

Anti-roll plate does not protrude far

Readjust correctly.

plate

enough beyond the knife edge.

 

 

 

 

Scraping noise during

Anti-roll plate protrudes too far beyond

Readjust correctly.

sectioning and specimen

the knife edge and is scraping against

 

return movement

the specimen.

 