9. Troubleshooting

 

 

 

Problem

Causes

Remedies

 

 

 

No or insufficient

Cooling system or electronic drive

Call technical service.

refrigeration of the

defective.

 

cryochamber

 

 

 

 

 

Sliding window collects

Air humidity and room temperature

Comply with the requirements

condensation

too high.

for the installation site.

 

 

 

No or insufficient

Cooling system or electronic drive

Call technical service.

refrigeration of the

defective.

 

specimen

 

 

 

 

 

Lamp does not work

Lamp defective.

Check lamp and replace it, if

 

 

necessary.

 

Switch defective.

Call technical service.

Both disinfection LEDs flash

UV radiation provided by UV tube no

alternately

longer sufficient.

Replace UV tube following the manufacturer’s instructions.

Image of an open-ended wrench

Contact technical service and

appears due to fault to be rectified.

follow the instructions given.