8. Working with the Instrument

Activate trimming mode.

Select the trimming thickness.

Switch on the VAC and select a low extraction level (between 1 and 2).

Start trimming by manually moving the handwheel or press the RUN/STOP and ENABLE buttons simultaneously to start motorized sectioning.

Use a brush to apply the section to a precooled specimen slide, then warm it from below using your finger.

Move the knife guard forward after removing the section.

Remove the knife or disposable blade (using the blade ejector!).

Insert the knife into the knife case.

8.4.3Cutting with Extraction – Anti-roll Guide installed

Shut off VAC (LED in VAC button goes out).

Switch from trimming to sectioning mode (important for section stretch- ing, as the valves work differently than in trimming mode.)

Set the desired section thickness.

Switch on the VAC and begin at level 1. If the section does not stretch correctly, turn the VAC knob higher in small increments.

Once the desired section is on the pressure plate, switch off the VAC! Carefully fold the anti-roll guide to the side and remove the section from the side.

After removing the section, wipe off moisture/condensate from the pressure plate – otherwise, the next sections will become jammed.

Cutting with extraction without anti-roll guide (brush technique) is not possible, as the position of the pressure plate means that no suitable air flow is obtained.

Leica CM1950 Cryostat

45

Page 45
Image 45
Leica CM1950 manual Cutting with Extraction Anti-roll Guide installed, Shut off VAC LED in VAC button goes out