6.Installation
Transport with a fork lift
• The instrument can be transported with a fork lift.
To ensure a safe transport with a fork lift 3 people are required: one operating the fork lift, and the other 2 holding the instrument on either side to prevent it from sliding down.
•At the installation site, unscrew the adjustable feet (1) with the open- end wrench No. 13 (see Fig. 12). This is absolutely necessary for a stable stand.
Total width with handwheel
Fig. 13
6.3Assembling the Handwheel
4
2
1
Fig. 14
Rotate the handwheel only if the refrigeration system is on and the cryochamber is cold.
•Insert the pin (1) of the handwheel shaft into the hole (2) of the handwheel.
•Tighten the screw (4) using the No. 6 Allen key.
•Place the protective cap on the screw (4). To dismount, proceed in reverse order.
22 | Operating Manual V1.1 – 05/2007 |