9. Troubleshooting

Problem

Causes

Remedies

 

Blunt knife.

Use different part of the knife.

 

Inappropriate section thickness.

Select correct section thickness.

 

Incorrect clearance angle.

Set correct angle.

 

Microtome not dried thoroughly

Dry microtome.

 

enough.

 

 

Dried specimen.

Prepare new specimen.

 

 

 

Tissue sticks to the anti-roll plate

Anti-roll plate is too warm or incorrectly positioned.

Fat on the corner or edge of anti-roll plate.

Anti-roll plate not correctly fixed. Rust on the knife.

Cool down anti-roll plate, or reposition correctly.

Remove fat from anti-roll plate.

Fix correctly.

Remove rust.

Flattened sections curl up when anti-roll plate is folded up

Anti-roll plate too warm.

Cool down anti-roll plate.

Sections tear

Temperature too low for the tissue cut.

Increase temperature and wait.

 

Blunt part, dirt, dust, frost or rust on

Remove cause.

 

the knife.

 

 

Top edge of the anti-roll plate

Replace anti-roll plate.

 

damaged.

 

 

Hard particles in the tissue.

- - -

 

Dirt on knife back.

Clean.

 

 

 

Cryostat non-operational

Power plug not properly connected.

 

Defective fuses, or circuit breaker has

 

triggered.

Check if properly connected. Replace fuses, or switch circuit breaker back on. If not possible, call technical service.

Specimen disc cannot be

Moisture on the underside cause the

removed

specimen to freeze to the freezing

 

shelf or specimen head.

Apply concentrated alcohol to the contact point.

50

Operating Manual V1.1 – 05/2007

Page 50
Image 50
Leica CM1950 manual Sections tear