8.Working with the Instrument

8.1Preparing Cutting Tools, Specimen Discs and Preparation Aids

The knives are extremely sharp! Handle with care!

Never try to catch a falling knife!

Place working materials such as the blade box or knives (in the knife case), brush, pincers or preparation needles and, where applicable, specimen discs into the cryostat chamber.

The necessary tools and preparation aids can be precooled on the (optional) movable shelf, making them available at all times in a conve- nient position for the user.

In addition, specimen discs can be precooled and stored in the storage system (see page 26, Fig. 21).

Fig. 36

For installation of knife/blade holder and installation in the chamber, refer to optional accessories on page 50).

8.2Turning the Instrument On

The instrument must be switched on at least 5 hours before the planned use.

The circuit breaker is used as power switch. The switch must be in the top position for switching on and in the bottom position for switching off. The switch must be accessible without obstruction.

• Close the sliding window.

Fig. 37

To avoid frost formation always put the cover on the quick freeze shelf.

Always cover the quick freeze shelf during breaks and overnight.