Leica CM1950 manual Cryochamber

Models: CM1950

1 74
Download 74 pages 62.9 Kb
Page 49
Image 49

 

 

9. Troubleshooting

 

 

 

Problem

Causes

Remedies

 

 

 

Ridged sections

Knife damaged.

Use different part of the knife.

 

Edge of anti-roll plate damaged.

Replace anti-roll plate.

 

 

 

Chatter during sectioning

Specimen insufficiently frozen onto

Refreeze specimen onto the disc.

 

the specimen disc.

 

 

Specimen disc not clamped tightly.

Check disc clamping.

 

Knife not clamped tightly enough.

Check knife clamping.

 

Specimen has been sectioned too

Refreeze specimen onto the disc.

 

thickly and has detached from disc.

 

 

Very hard, inhomogenous specimen.

Increase section thickness; reduce

 

 

specimen surface area if ecessary.

 

Blunt knife.

Use different part of the knife.

 

Knife profile inappropriate for

Use knife with different profile.

 

specimen cut.

 

 

Incorrect clearance angle.

Set correct angle.

 

 

 

Condensation on anti-roll plate and knife during cleaning

Brush, forceps and/or cloth too warm.

Store all tools on storage shelf in

 

the cryochamber.

Anti-roll plate damaged

Anti-roll plate too high above the knife

Replace anti-roll plate.

after adjustment

edge. The adjustment was carried out

Be more careful next time!

 

in the direction of the knife.

 

 

 

 

Thick-thin sections

Temperature incorrect for the

Select correct temperature.

 

tissue cut.

 

 

Knife profile inappropriate for the

Use knife with different profile

 

specimen cut.

(c or d).

 

Ice buildup on the knife back.

Remove ice.

 

Handwheel speed not uniform.

Adapt speed.

 

Knife not clamped tightly enough.

Check knife clamping.

 

Specimen disc not clamped tightly

Check disc clamping.

 

enough.

 

 

Cryocompound applied to cold

Apply cryocompound to warm disc,

 

specimen disc; specimen detached

mount specimen and freeze.

 

from disc after freezing.

 

Leica CM1950 Cryostat

49

Page 49
Image 49
Leica CM1950 manual Cryochamber