ENTER

ENABLE/DISABLE

SET POLARITY

 

 

 

 

 

CLEAR

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FIGURE THIRTEEN: Analyzer’s INPUT Menu

6.2.4. Select ENABLE/DISABLE. Press Enter.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DIGITAL INPUT CONTROL

 

 

 

 

 

 

 

 

CO

NTROL I

S: ON

 

 

 

 

 

ENTER

 

 

 

 

 

 

 

 

 

CHANGE TO:

OFF

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ARROWS TO CHANGE

 

 

 

 

 

CLEAR

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ENT = SET

 

CLR = EXIT

 

 

 

 

 

 

 

FIGURE FOURTEEN: DIGITAL INPUT CONTROL screen

6.2.5.The digital input is either in a state of ON or OFF. To change this state, press the up/down arrows to toggle between the two choices. When the state in the change is correct, press Enter to set. To allow input, the state must be ON. Press Clear to return to the INPUT Menu.

ENABLE/DISABLE

SET POLARITY

ENTER

CLEAR

FIGURE FIFTEEN: Analyzer’s INPUT Menu

6.2.6. Select SET POLARITY by using the up/down arrows, press Enter.

DIGITAL INP. POLARITY RUN WHEN: OPEN CHANGE TO: CLOSED

ARROWS TO CHANGE

ENT = SET

CLR = EXIT

ENTER

CLEAR

FIGURE SIXTEEN: DIGITAL INP. POLARITY Screen

6.2.7.The DIGITAL INP. POLARITY is either in a OPEN or CLOSED condition. To change this condition, press the up/down arrows to toggle between the two conditions. When the condition is correct, press Enter to set. It is critical to

have the polarity set to the same condition as the switch, whether it is

GE Analytical Instruments, Inc. 2005H

Page 15 of 16

DIN 30001 Rev. E

 

 

TOC,FLOW SWITCH UPGRADE 2.12CBI