CAUTION:
IF THERE ARE ANY VARIATIONS, CONSTANTS MUST BE ENTERED THROUGH THE FACTORY ACCESS MENU. ACCESS TO THIS MENU IS LIMITED AND WILL REQUIRE A PASSWORD. ENSURE THAT ACCESS TO THIS MENU IS RESTRICTED. ANY WRONGFUL DATA ENTRY WILL RESULT IN THE ANALYZER PERFORMING OUT OF SPECIFICATION!!!
5.6.1.1.From the MAIN menu select SETUP, press Enter.
5.6.1.2.From the SETUP menu, select CALIBRATE, press Enter.
5.6.1.3.Select TC ZERO OFFSET, press Enter.
TC ZERO CORRECTION
ENT=SAVE CHAR/NUM
CLR =RESTART/EXIT
ARROWS =CHANGE CHAR
ENTER
CLEAR
FIGURE SEVEN: TOC ZERO OFFSET Screen
5.6.1.4.Enter 911 by using the arrows, and then press Enter.
NOTE:
ENTERING 911 WILL IN NO WAY AFFECT THE VALUE OF THE TC ZERO
CORRECTION.
FACTORYACCESSV. 0.00
|
| ERASECARD |
ERASEERRS |
| ANALOG # |
PRNTSETUP |
| FACTMODES |
SETEEPROM | RAW DATA | |
COND ZERO |
|
ENTER
CLEAR
FIGURE EIGHT: FACTORY ACCESS Menu
5.6.1.5.Using the arrows, select SETEEPROM, and press Enter.
5.6.1.6.Constants will be displayed in the same order as the printed or recorded constants list.
5.6.1.7.Once scrolling through the list has begun, the entire list must be seen. DO NOT PRESS THE CLEAR KEY. Pressing the Clear key will result in the erasing of the Constant Value entered. To move up or down the list use the arrow keys.
C1 CONST:
0.800600
ENT = CHANGEVALUE CLR= SET TODEFAULT ARROWS= NEXT VALUE
ENTER
CLEAR
GE Analytical Instruments, Inc. ♥2005H | Page 8 of 16 | DIN 30001 Rev. E |
|
| TOC,FLOW SWITCH UPGRADE 2.12CBI |