CONSTANT VALUE

CONSTANT VALUE AFTER

CONSTANT VALUE PRIOR

 

NAME

UPGRADE

TO UPGRADE

 

 

 

 

 

ANALOG MAX:

Same as prior value

 

 

TOTAL ACID VOL:

150

 

 

TOTAL OXID VOL:

150

 

 

ALARM #1 TYPE:

Same as prior value

 

 

ALARM #2 TYPE:

Same as prior value

 

 

ALARM #1 LEVEL:

Same as prior value

 

 

ALARM #2 LEVEL:

Same as prior value

 

 

LAMP MAX DAYS:

182

 

 

HOSE MAX DAYS:

365

 

 

OXID MAX DAYS:

90

 

 

HYSTERESIS CNT:

40

 

 

TC/IC OFFSET (60Hz):

Same as prior value

 

 

SLOPE PPM/HOUR:

5000

 

 

S1 TOC ACID FLOW:

0.2

 

 

S1 TOC OXID FLOW:

0

 

 

S2 TOC ACID FLOW:

0.4

 

 

S2 TOC OXID FLOW:

0.2

 

 

S3 TOC ACID FLOW:

0.75

 

 

S3 TOC OXID FLOW:

0.5

 

 

S4 TOC ACID FLOW:

0.75

 

 

S4 TOC OXID FLOW:

2.5 or 2.0

 

 

 

same as prior value

 

 

REFILL DELAY:

15

 

 

SERIAL NUMBER:

Same as prior value

 

 

TC TEMP CORR A:

2.03149

 

 

TC TEMP CORR B:

-0.0593627

 

 

TC TEMP CORR C:

0.00101571

 

 

TC TEMP CORR D:

-5.7161E-06

 

 

IC TEMP CORR A:

2.57516

 

 

IC TEMP CORR B:

-0.0975902

 

 

IC TEMP CORR C:

0.00194737

 

 

IC TEMP CORR D:

-1.34084E-05

 

 

DAC RANGE MIN:

Same as prior value

 

 

DAC RANGE MAX:

Same as prior value

 

 

PRINTOUT INTERVAL:

0

 

 

SAMPLE STEPZERO:

100

 

 

SAMPLE FLOWRATE:

55 or 71

 

 

 

same as prior value

 

 

SYRINGE MAX POINT:

31000

 

 

SYRINGE MAX RANGE:

3100

 

 

SYRINGE ZERO RANGE:

150

 

 

REAGENT FLUSH REPS:

2

 

 

NUMBER OF STREAMS:

4

 

 

AC FREQ. SETTING:

Same as prior value

 

 

 

 

 

 

GE Analytical Instruments, Inc. 2005H

Page 12 of 16

DIN 30001 Rev. E

 

 

TOC,FLOW SWITCH UPGRADE 2.12CBI