CONSTANT VALUE | CONSTANT VALUE AFTER | CONSTANT VALUE PRIOR |
|
NAME | UPGRADE | TO UPGRADE |
|
|
|
|
|
ANALOG MAX: | Same as prior value |
|
|
TOTAL ACID VOL: | 150 |
|
|
TOTAL OXID VOL: | 150 |
|
|
ALARM #1 TYPE: | Same as prior value |
|
|
ALARM #2 TYPE: | Same as prior value |
|
|
ALARM #1 LEVEL: | Same as prior value |
|
|
ALARM #2 LEVEL: | Same as prior value |
|
|
LAMP MAX DAYS: | 182 |
|
|
HOSE MAX DAYS: | 365 |
|
|
OXID MAX DAYS: | 90 |
|
|
HYSTERESIS CNT: | 40 |
|
|
TC/IC OFFSET (60Hz): | Same as prior value |
|
|
SLOPE PPM/HOUR: | 5000 |
|
|
S1 TOC ACID FLOW: | 0.2 |
|
|
S1 TOC OXID FLOW: | 0 |
|
|
S2 TOC ACID FLOW: | 0.4 |
|
|
S2 TOC OXID FLOW: | 0.2 |
|
|
S3 TOC ACID FLOW: | 0.75 |
|
|
S3 TOC OXID FLOW: | 0.5 |
|
|
S4 TOC ACID FLOW: | 0.75 |
|
|
S4 TOC OXID FLOW: | 2.5 or 2.0 |
|
|
| same as prior value |
|
|
REFILL DELAY: | 15 |
|
|
SERIAL NUMBER: | Same as prior value |
|
|
TC TEMP CORR A: | 2.03149 |
|
|
TC TEMP CORR B: |
|
| |
TC TEMP CORR C: | 0.00101571 |
|
|
TC TEMP CORR D: |
|
| |
IC TEMP CORR A: | 2.57516 |
|
|
IC TEMP CORR B: |
|
| |
IC TEMP CORR C: | 0.00194737 |
|
|
IC TEMP CORR D: |
|
| |
DAC RANGE MIN: | Same as prior value |
|
|
DAC RANGE MAX: | Same as prior value |
|
|
PRINTOUT INTERVAL: | 0 |
|
|
SAMPLE STEPZERO: | 100 |
|
|
SAMPLE FLOWRATE: | 55 or 71 |
|
|
| same as prior value |
|
|
SYRINGE MAX POINT: | 31000 |
|
|
SYRINGE MAX RANGE: | 3100 |
|
|
SYRINGE ZERO RANGE: | 150 |
|
|
REAGENT FLUSH REPS: | 2 |
|
|
NUMBER OF STREAMS: | 4 |
|
|
AC FREQ. SETTING: | Same as prior value |
|
|
|
|
|
|
GE Analytical Instruments, Inc. ♥2005H | Page 12 of 16 | DIN 30001 Rev. E | |
|
| TOC,FLOW SWITCH UPGRADE 2.12CBI |