5.2.9.Select each consumable, press Enter. Write the information on the Printout or the Constants Record Form. Record date and percent left for the OXIDIZER. Record the percent left for the LAMP, PUMP TUBING and ACID.

5.2.10.REAGENT FLUSH is not a consumable to record.5.2.11.Print or record Errors and Warnings if desired.

5.2.12.Data on the Ram Card will not be erased or affected in any way upon performing this procedure.

5.3.Replace the EPROM chip (IC chip).CAUTION:

Prior to performing the following instructions it is essential, to be grounded according to established procedures. Failure to be properly grounded will result in damage to the EPROM Chip and void the warranty.

5.3.1.Turn off the analyzer power and disconnect the power cord.

5.3.2.Lift the analyzer cover and locate the PC electronics board, refer to Figure Six.

5.3.3.Depending on the type of chip holder in the analyzer, different techniques are required to replace the existing EPROM with the EPROM labeled 2.12CBI. Facing the analyzer’s front panel, locate the EPROM socket on the left side of the PC electronics board near the rear of the analyzer. If the chip socket has a metal lever at the top, it is a Zero Insertion Force Socket (Step 5.3.4). If the chip socket does not have a lever, it is a Low Profile Socket (step 5.3.5). Figure Six illustrates the low profile socket. Use the appropriate instruction to replace the chip.

IC SOCKET

(POWER SUPPLY)

PC BOARD

 

FRONT OF TOC

 

PC BOARD

 

RIGHT INDEX

SCREW DRIVER HERE

FINGER HERE

SOCKET

UPPER TAB

 

 

 

CARRIER

EPROM

LOWER TAB

NOTCH

FIGURE SIX: Low Profile Socket Replacement

GE Analytical Instruments, Inc. 2005H

Page 5 of 16

DIN 30001 Rev. E

 

 

TOC,FLOW SWITCH UPGRADE 2.12CBI