TC ZERO CORRECTION

 

 

 

 

 

 

 

 

 

 

 

-1.3250000

 

 

 

 

 

 

ENTER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ENT = SAVE CHAR/NUM

 

 

 

 

 

 

 

 

 

 

 

 

CLEAR

 

 

CLR = RESTART/EXIT

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ARROWS = CHANGE CHAR

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FIGURE THREE: TOC ZERO OFFSET Screen

5.2.3.Enter 911 by using the arrows, and then press Enter.

ENTERING 911 WILL IN, NO WAY AFFECT THE VALUE OF THE TC ZERO CORRECTION.

 

 

 

 

 

 

 

 

 

 

 

FACTORYACCESSV. 0.00

 

 

 

 

 

 

 

 

 

SETSYRNGE

 

 

 

 

 

ENTER

 

 

 

 

 

 

 

 

 

ERASEERRS

SETSERVO

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PRNTSETUP

DUMPRAM

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETEEPROM

SET4-20MA

 

 

 

 

 

CLEAR

 

SYRNGEMOV

SYRNGPULS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FIGURE FOUR: FACTORY ACCESS Menu

5.2.4.Using the arrows, select SETEEPROM, and press Enter.

5.2.5.Constants will be displayed in the sequence listed on the Constants Record Form. Record each constant value on the form contained within this document.

IMPORTANT:

It is critical that this Printout or Constants Record Form is retained for comparison to the Constant Printout or Record Form at end of upgrade procedure. These Constants will have to be checked and any variations must be corrected after the upgrade is completed.

5.2.7.Record Dates and Percents for consumables on the Constants Record Form. These will need to be reentered after upgrade.

5.2.8.From the MAIN MENU select CONSUMABLES, press Enter.

LAMP OXIDIZER

ACID PUMPTUBING

REAGENTFLUSH

ENTER

CLEAR

FIGURE FIVE: Consumables Menu

GE Analytical Instruments, Inc. 2005H

Page 4 of 16

DIN 30001 Rev. E

 

 

TOC,FLOW SWITCH UPGRADE 2.12CBI