| TC ZERO CORRECTION |
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
| ENTER | |
|
|
|
|
|
|
|
|
|
|
| ||
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
| |
|
| ENT = SAVE CHAR/NUM |
|
|
|
|
|
| ||||
|
|
|
|
|
| CLEAR | ||||||
|
| CLR = RESTART/EXIT |
|
|
|
|
| |||||
|
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
| |||||
ARROWS = CHANGE CHAR |
|
|
|
|
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
FIGURE THREE: TOC ZERO OFFSET Screen
5.2.3.Enter 911 by using the arrows, and then press Enter.
ENTERING 911 WILL IN, NO WAY AFFECT THE VALUE OF THE TC ZERO CORRECTION.
|
|
|
|
|
|
|
|
|
|
| FACTORYACCESSV. 0.00 |
|
|
|
|
|
| ||
|
|
| SETSYRNGE |
|
|
|
|
| ENTER |
|
|
|
|
|
|
|
| ||
| ERASEERRS | SETSERVO |
|
|
|
|
|
| |
|
|
|
|
|
|
| |||
|
|
|
|
|
|
| |||
| PRNTSETUP | DUMPRAM |
|
|
|
|
|
| |
|
|
|
|
|
|
| |||
| SETEEPROM |
|
|
|
|
|
| CLEAR | |
| SYRNGEMOV | SYRNGPULS |
|
|
|
|
| ||
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
FIGURE FOUR: FACTORY ACCESS Menu
5.2.4.Using the arrows, select SETEEPROM, and press Enter.
5.2.5.Constants will be displayed in the sequence listed on the Constants Record Form. Record each constant value on the form contained within this document.
IMPORTANT:
It is critical that this Printout or Constants Record Form is retained for comparison to the Constant Printout or Record Form at end of upgrade procedure. These Constants will have to be checked and any variations must be corrected after the upgrade is completed.
5.2.7.Record Dates and Percents for consumables on the Constants Record Form. These will need to be reentered after upgrade.
5.2.8.From the MAIN MENU select CONSUMABLES, press Enter.
LAMP OXIDIZER
ACID PUMPTUBING
REAGENTFLUSH
ENTER
CLEAR
FIGURE FIVE: Consumables Menu
GE Analytical Instruments, Inc. ♥2005H | Page 4 of 16 | DIN 30001 Rev. E |
|
| TOC,FLOW SWITCH UPGRADE 2.12CBI |