FIGURE NINE: First Constant Screen

5.6.1.8.Use the Constants Record Form as a guide. If a number is on the Constants Record Form, the new constant value must be that value. If the Constants Record Form states “same as prior value”, change the value to match. If the constants Record Form states “do not change”, do not change the value.

5.6.1.9.If the value displayed does not have to be changed, press the UP arrow key to progress to the next constant value on the list.

5.6.1.10.If the value displayed must be changed, press Enter. Use the arrows to scroll to the first digit/decimal/minus sign, press Enter to save that character and go to the next.

5.6.1.11.When upgrading Model 810 and Model 820 from very early Firmware revisions, it is critical to check that the constant setting for AC FREQUENCY is set to 50. Model 800 must be set to 60. There are some 100 volt analyzers that are 60Hz, in this case use the C1 XFER. C2 Xfer and TC/IC OFFSET of the appropriate frequency. The 50 Hz calibration constants are toward the bottom of the list.

5.6.1.12.Press Enter to save the constant value. Press the Up arrow key to progress to the next constant value. Press the Down Arrow to return to previous constant value. Press Clear to return the constant to its default value.

5.6.1.13.There may be new constants when upgrading, make all changes as instructed on the Constants Record Form to keep the analyzer in specification.

5.6.1.14.When the last constant value (LINERARITY VAR.E) has been verified, the following screen will appear:

END OF EEPROM LIST EXIT? ENT=Y CLR =N

ENTER

CLEAR

FIGURE TEN: Set EEPROM Exit screen
5.6.1.15.Pressing Enter will return to the MAIN Menu.

5.6.1.16.Pressing Clear will return to the last value of the constant list. Scroll using the DOWN arrow to change any constant value. Scroll back to the end of the list to exit. Press Enter to return to the MAIN Menu.

5.7.From the MAIN Menu, a final configuration must be performed. Before the analyzer may be used, the flowrates and the sampling mode must be reset. Refer to the user manual for more a detailed procedure.

5.7.1.Set Clock – From the MAIN Menu, select SETUP. Press Enter. Select CLOCK. Press Enter. Use the arrow buttons to scroll to each digit option. Press Enter

GE Analytical Instruments, Inc. 2005H

Page 9 of 16

DIN 30001 Rev. E

 

 

TOC,FLOW SWITCH UPGRADE 2.12CBI