6. Firmware Configuration

It is necessary to configure the binary inputs following the firmware upgrade. Follow these instructions to configure inputs and select proper polarity.

If more information is required read the Operation and Maintenance Manual concerning setting up Binary/Digital Inputs. Thoroughly read the chapters on Installation and Operation.

The purpose of the Binary/Digital Input Module is to send a binary/digital signal to the analyzer that will change operating mode to a PAUSED state.

6.1.Wire the analyzer for binary/digital inputs:

6.1.1.Connect the flow switch to the provided terminal block on the Binary Switch Module located on the Auxiliary Port, which is on the left side of the analyzer.

6.1.2.Use a continuous closure contact system only. Applying voltage to the Binary Switch Module may damage the analyzer.

6.1.3.Use either opening in the Binary Switch Module. There is no polarity on the terminal block.

6.2.Setup the analyzer for binary/digital input:

6.2.1.From the MAIN Menu select SETUP using the up/down arrows.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

START TOC

 

 

 

 

 

 

 

 

 

SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ENTER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HISTORY

 

CALIBRATE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ERRORS

CONSUMABLES

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CLEAR

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FIGURE ELEVEN: Analyzer’s MAIN Menu

6.2.2. Press Enter.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CLOCK

 

PRINTER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ENTER

 

 

REAGENTS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

INPUTS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PRESETS

OUTPUTS

 

 

 

 

 

 

 

 

 

 

 

 

 

CLEAR

 

 

 

 

 

 

 

 

 

SAMPLE MODE

PW

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

UV LAMP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FIGURE TWELVE: Analyzer’s SETUP Menu

 

 

 

 

 

 

6.2.3. Select INPUTS by using the up/down arrows. Press Enter.

GE Analytical Instruments, Inc. 2005H

Page 14 of 16

DIN 30001 Rev. E

 

 

TOC,FLOW SWITCH UPGRADE 2.12CBI