to save digits. After resetting the clock, Warning 35 will occur until the power has been cycled two times.

5.7.2.Set Flowrates – The Firmware allows selection of different flow rates for on- line sampling and grab (container) sampling. The grab flow rates can be viewed and changed from the MAIN – SETUP/REAGENTS menu and from the GRAB SAMPLING – SETUP/REAGENTS menu. Make sure to view and change the desired flow rates for the sample mode being used.

5.7.3.From the SETUP menu, choose REAGENTS. From the REAGENTS menu, choose ONLINE FLOW RATES. Set both reagent flowrates to the desired level. Select the sampling mode.

5.7.3.From the SETUP menu, choose SAMPLE MODE.

5.7.3.1.Select ON LINE SAMPLING or GRAB SAMPLING. If GRAB SAMPLING is selected, START TOC will access the GRAB SAMPLING menu, where additional setup options may be implemented.

5.7.4.Select Frequency, only if AC FREQ. SETTING” is set to 50 for the 810 and 820 analyzers.

5.7.5.1.From SETUP, select 50/60HZ.5.7.5.2.Select the appropriate Frequency for the analyzer.5.7.6.Reset Acid/Oxidizer/UV Lamp/Pump Tubing dates and percentages.5.7.6.1.Select CONSUMABLES from the MAIN Menu, press Enter.5.7.6.2.Select ACID, OXIDIZER, UV LAMP, or PUMP TUBING, press Enter.

5.7.6.3.For the oxidizer, select SET % LEFT and enter the percent recorded on the printout or the Constants Record Form. Use the arrow keys to scroll through digits and press the Enter key to set the digits. Select SET INSTALL DATE, press Enter. Use the arrow keys to scroll through digits and press the Enter key to set the digits.

5.7.6.4.For the ACID, LAMP and PUMP TUBING, select SET % LEFT and enter the percent recorded on the printout or the Constants Record Form. Use the arrow keys to scroll through digits and press the Enter key to set the digits.