2. General Description

Rear panel controls

Fig. 6. Control unit. Rear panel.

No.

Item

Description

 

17

Mains voltage selector

Selects mains voltage 110, 130, 220 and 240 V

18

Fuse holder

Mains fuses:

1 x 250 mA for 110-130 V 60 Hz

 

 

 

1 x 125 mA for 220-240 V 50 Hz

19

Signal outputs

The output signal is a 10 mV DC signal. The

 

 

terminals are connected to a potentiometric

 

 

recorder via shielded cables supplied. The

 

 

UV-1 is normally earthed via the mains ground

20

Mains inlet

Inlet for the mains cable

21

Optical unit inlet

Inlet for the optical unit

22

Ground terminal

To connect the shield of an output signal cable

 

 

 

 

Note: Do not attempt to ground the control unit via the recorder ground as the UV-1 is grounded via the mains ground.

10