3. Installation
Monitor
Product | Material of | Path | Total dead | Illuminated | Pressure | Application |
Code No. | wetted parts | length | volume | volume | limit | area |
|
|
|
|
|
|
|
2 mm | 80 µl | 2 µl | 0.3 MPa | Standard | ||
optical quartz |
|
|
| (3 bar) | Chromatography | |
10 mm | 24 µl | 8.7 µl | 1.0 Mpa | FPLC, Standard | ||
optical quartz,titanium |
|
|
| (10 bar) | Chromatography | |
3 mm | 3 mm | 50 µl | 3 µl | 1.0 MPa | Preparative FPLC, | |
optical quartz |
|
|
| (10 bar) | Standard Chromatography | |
10 mm | 10 mm | 250 µl | 8.7 µl | 1.0 MPa | Standard | |
optical quartz |
|
|
| (10 bar) | Chromatography | |
5 mm | Silicon rubber, | 5 mm |
|
| 0.2 MPa* | Industrial scale |
Industrial | polypropylene, |
|
|
| (2 bar) |
|
optical quartz |
|
|
|
|
|
* Flow rate 300 I/h at 0.1 MPa
All flow cell are mounted in the holders, ready to be installed directly into the cell housing. Release the locking knob (Fig. 4:6) by turning it against the direction of the arrow and insert the flow cell in its holder into the optical unit. Lock it in position by turning the locking knob fully in the direction of the arrow.
For more information please refer to the instruction sheet supplied with the respectively flow cell.
Fig. 7.
Fig. 8.
Fig. 9. 3 mm flow cell | Fig. 10. 10 mm flow cell | Fig. 11. Industrial flow cell |