6.
6. Trouble-shooting
The
Symptom | Cause | Remedy | ||
Pilot light does not light | 1. Mains cord not plugged in | Check that mains cord is plugged in | ||
| 2. | Fuse blown | Replace fuse. If fuse blows again | |
|
|
|
| immediately consult GE Healthcare representative |
| 3. No voltage at mains socket | Check by plugging table lamp in | ||
|
|
| ||
Pilot light on, no recorder 1. | Optical unit not connected | Check that connecting cable is | ||
response |
|
| to control unit | plugged in |
| 2. | Control unit not connected | Check connection between output | |
|
|
| to recorder | terminals and recorder |
| 3. | Recorder not operating | Check recorder function | |
| 4. | Recorder zero not set correctly | Zero recorder | |
| 5. | Recorder range incorrect | Set recorder range to 10 mV | |
| 6. | Wrong filter | Check that the filter corresponds to | |
|
|
|
| converter or aperture |
| 7. | Filter not pushed in fully | Push in filter | |
| 8. | Shutter in place | Remove shutter and insert converter | |
|
|
|
| or aperture as appropriate |
|
|
|
|
|
Excessive noise | 1. |
| Poor ground contact | Check contact to ground |
| 2. |
| Excess noise on mains supply | Use alternative power source or |
|
|
|
| remove source of disturbance |
| 3. | Recorder connections | Check connection between | |
|
|
| incorrect | output terminals and recorder |
| 4. |
| Recorder range incorrect | Set recorder on 10 mV range |
| 5. |
| Dirty cell | Clean cell |
| 6. |
| Deposits on optical surfaces | Clean filter and converter |
|
|
|
| Relocate optical unit in a clean environment |
| 7. | Solvent with high UV absorption | Change to a more suitable solvent | |
| 8. | Lamp not warmed up | 2 h | |
| 8. | Bubbles passing through the cell | ||
| 9. |
| Aging lamp | Check lamp and replace if necessary |
|
|
|
|
|
Excessive baseline drift | 1. |
| Variable absorbance gradient | Compensate by use of reference cell |
| 2. |
| Contaminated solvent | Use fresh solvent. Check that plastic tubing |
|
|
|
| does not leak UV absorbing substances |
| 3. |
| Large variation in ambient | Relocate optical unit or remove |
|
|
| temperature | source of temperature change |
| 4. |
| Instrument | Allow 2 hours |
| 5. | Condensation forming in | Flush reference cell with dry gas or | |
|
|
| empty reference | fill it with the appropriate solvent |
|
|
| ||
Long term noise, | 1. Variations in ambient temperature | Relocate optical unit or protect | ||
often regular waves in |
|
| especially in cold room | from draught |
recorder reponse | 2. Flow rate variations | Check pump system and column packing | ||
| 3. Poor ground contact | Check contact to ground | ||
| 4. Bubbles passing cell | |||
| 5. | Dirty cell | Clean cell | |
| 6. Deposits on optical surfaces | Clean filter and converter | ||
|
|
|
| Relocate optical unit in a clean environment |
|
|
|
|
|
23