4. Operation
4.3Conversion table T% to AU and OD
4.4 Start-up
4.5Stabilization time
T% | AU | OD | OD |
|
| (3 mm cell) | (10 mm cell) |
|
|
|
|
1 | 2.000 | 6.667 | 2.000 |
5 | 1.301 | 4.337 | 1.301 |
10 | 1.000 | 3.333 | 1.000 |
15 | 0.824 | 2.747 | 0.824 |
20 | 0.699 | 2.330 | 0.699 |
25 | 0.602 | 2.007 | 0.602 |
30 | 0.523 | 1.743 | 0.523 |
32 | 0.500 | 1.667 | 0.500 |
35 | 0.456 | 1.520 | 0.456 |
40 | 0.398 | 1.327 | 0.398 |
45 | 0.347 | 1.157 | 0.347 |
50 | 0.301 | 1.003 | 0.301 |
55 | 0.260 | 0.867 | 0.260 |
60 | 0.222 | 0.740 | 0.222 |
63 | 0.200 | 0.667 | 0.200 |
65 | 0.186 | 0.620 | 0.186 |
70 | 0.155 | 0.517 | 0.155 |
75 | 0.125 | 0.417 | 0.125 |
79 | 0.100 | 0.333 | 0.100 |
80 | 0.097 | 0.323 | 0.097 |
85 | 0.070 | 0.233 | 0.070 |
89 | 0.050 | 0.167 | 0.050 |
90 | 0.046 | 0.153 | 0.046 |
95 | 0.022 | 0.073 | 0.022 |
95.5 | 0.020 | 0.067 | 0.020 |
97.7 | 0.010 | 0.033 | 0.010 |
Note: Always ensure that all liquid passing through the flow cell are degassed to prevent any air bubble formation in the cell. Liquids must be filtered to remove any particulate material and prevent any blockage.
1.Check that the monitor is correctly installed and that the sample cell is filled with the appropriate eluent. The reference cell may be left closed with air in the cell unless the
2.If the instrument has been switched off, turn on the mains switch (Fig. 5:16) and refer to Section 4.5, to restabilize the instrument.
3.Check that the appropriate filter and converter or aperture are in place and fully inserted.
At normal laboratory temperatures the
For coldroom operation below 10 °C, install
16