2. General Description
2.3 Control unit | Front panel controls |
12
13
14
15
16
Fig. 5. Control unit. Front panel.
No. | Item | Description |
|
|
|
12 | Absorbance range | Selector for the desired absorbance range |
| selector | 0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1 or 2 AU full scale |
|
| deflection. In the position SHORT, the signal output |
|
| terminals are disconnected from the rest of the control |
|
| circuitry and |
|
| setting zero on the recorder. |
13 | Mode switch | The AU / % T switch, enables the |
| AU or %T | to monitor the absorbance or transmission of a flowing |
|
| liquid. When the |
|
| the other position corresponds to the optical |
|
| absorbance which will give a full scale deflection (fsd) |
|
| on a recorder with a sensitivity of 10 mV. Absorbances |
|
| are measured relative to air or a liquid which is present |
|
| in the reference cell |
|
| When the recorder is used together with |
|
| % T mode, the absorbance range selector switch is |
|
| used, together with baseline adjust, to adjust the |
|
| recorder response corresponding to 100 % T |
14 | Baseline adjust | |
15 | Indicator lamp | Indicates power is on |
16 | Mains switch | The |
|
| (upper) position. The indicator lamp will light to show |
|
| that the mains voltage is on |
|
|
|