7.Technical Specifications

Wavelength range Lamp

Filters

Operating modes Full scale ranges Noise

Linearity

Temperature drift

Long term drift

Time constant

Recorder output

Environment

Power consumption

Power supply, voltage frequency

Dimensions (LxWxD)

Weight

254 nm, 280 nm and 405 nm

Hg lamp: for 254, 280, and 405 nm

Life time lamp: 8000 hours converter: 2000 hours

Interference. Stray light maximum 0,1 % at 254 nm, maximum 0.8% at 280 nm

AU or Transmission

0.01, 0.02, 0.05, 0.1, 0.2, 0.5, 1 or 2 AUFS

4 x 10-5AU peak to peak maximum at

254 nm (dry cell)

2 x 10-4 AU peak to peak (typical at 254 nm in flowing liquid)

5x 10-4 AU peak to peak (typical at 280 nm flowing liquid)

At 254 nm, better than ± 3% to 2 AU At 280 nm, better than ± 5% to 1 AU

2 x 10-4 AU/°C typical with dry cell at

254 nm <2 x 10-3 AU/°C typical with dry cell at 280nm

1 x 10-4 AU/h at 254 nm, constant

temperature after 2 hours warm-up 4 x 10-4 AU/h at 280 nm, constant

temperature after 2 hours warm-up 1.5 s to 90% of FSD at all ranges 0-10 mV

0to +40 °C, 20-95% relative humidity,

84-106 kPa (840-1060 mbar) atmospheric pressure

20 VA

100/120/220-230/240 V- 50-60 Hz

Control unit: 180x145x75 mm

Optical unit: 180x145x75 mm

Control unit: 1.6 kg Optical unit: 1.7 kg

EMC standards

This product meets the requirement of the EMC Directive 89/336/EEC through the harmonized standards

 

EN 50081-2 (emission) and EN 50082-1 (immunity)

 

Note: This is a class A product. In a domestic environment this product may cause radio interference in which

 

case the user may be required to take adequate actions.

 

Note: The declaration of conformity is valid for the instrument when it is:

 

used in laboratory locations

 

used in the same state as it was delivered from GE Healthcare Bio-Sciences AB except for alteration described in the

 

User Manual

 

used as “stand alone” unit or connected to other CE labelledGE Healthcare products or other products as

 

recommended.

Safety standards

This product meets the requirement of the Low Voltage Directive (LVD) 73/23/EEC through the harmonized

 

standard EN 61010-1.

24