2. General Description
2. General Description
2.1 Basic principle | GE Healthcare LKB Monitor | |
| containing the flow cell, lamp, filter assemblies and preamplifiers, and a | |
| control unit containing the signal processing circuits. The two units are | |
| connected via a | |
| the optical unit. Connection to the recorder and the mains supply is | |
| made via the control unit. | |
| The single path Monitor | |
| sample cell and a reference cell. Optical path lengths are 10 mm, | |
| 3 mm, or 2 mm depending upon the flow cell chosen. | |
| Note: | Flow cell |
The sample cell and reference cell are angled to receive light from the same point in the lamp (Fig. 1) thus assuring a stable
The lamp is powered by a stabilized DC to AC converter operating at 20 kHz to maximize its efficiency. Lamp output is independent of variations in the line voltage.
Light from the lamp passes through an aperture (254 nm, 405 nm) or a fluorescence converter (280 nm) and through the reference cell and sample cell. The light output from each cell passes through an interference filter and falls onto a solid state
Fig. 1. Optical path.