5. Maintenance
5.1General precautions
5.2Cleaning the flow cell
5.Maintenance
To ensure trouble free running, users are advised to observe the following precautions:
All liquids passing through the flow cell should be free of suspended particles.
All liquids should be degassed to prevent air bubble formation in the flow cell.
Never allow buffer solutions to dry out in the cell. Always rinse the flow cell thoroughly with distilled water after use.
Handle interference filters with care. Never touch the optical surfaces or expose them to temperatures above 60 °C.
For trouble free operation of the
The cells may be inspected for particles by removing the cell holder and examining the light paths with a magnifying glass.
If the cell contains trapped particles proceed as follows:
1.Remove the cell holder from the optical unit.
2.Connect a syringe to the outlet tubing and squirt a clean solution of ethanol in distilled water (50% v/v) through the cell in small aliquots. Examine the cell from time to time to see when the particles have been washed out.
3.Rinse the cell with
4.Reconnect the cell holder to the system to be monitored.
Most
Oily deposits, which increase the tendency to trap bubbles, can be removed by rinsing the flow cell first with a
18