ADDENDUM
I. Generally Applicable Annotations
Consistent with good engineering practice, leads attached to customer accessible signal/monitoring ports (such as the
II. CE'92 Product Specific Annotations
When tested for radiated susceptibility as called for in EN
6621A
When subjected to radiated field strengths of 3 volts/meter in the vicinity of 90 MHz, the full scale programming accuracy of channel 2 increases from 31 millivolts at 20 volts output to 700 millivolts.
The accuracy reverts to the published value of 31 millivolts when the field is reduced to 2 volts/ meter.
When subjected to radiate field strengths of 3 volts/meter in the vicinity of 200 MHz, the full scale readback accuracy of channel 1 increases from 23 millivolts at 5 volts output to 100 millivolts.
The accuracy reverts to the published value of 30 millivolts when the external field is reduced to 2 volts/meter.
6623A
When subjected to radiated field strengths of 3 volts/meter in the vicinity of 160 MHz, the full scale programming accuracy of channel 1 increases from 31 millivolts at 20 volts output to 50 millivolts.
The accuracy reverts to the published value of 31 millivolts when the field is reduced to 2.5 volts/ meter.
When subjected to radiate field strengths of 3 volts/meter in the vicinity of 160 MHz. the full scale readback accuracy of channel 1 increases from 30 millivolts at 5 volts output to 50 millivolts.
The accuracy reverts to the published value of 30 millivolts when the external field is reduced to 2.5 volts/meter.
114 Addendum