SECTION 6—TROUBLESHOOTING
SYMPTOMS
FAULTS
SOLUTION
Patient Lift feels loose. | Mast/Base joint loose. | Refer to Replace the locknut. on page 26. |
| Tie - Rods are loose. |
|
|
|
|
Noisy or dry sound from pivots. | Needs lubrication. | Refer to Lubricating the Lift on page 22. |
|
|
|
Electric actuator fails to lift when button |
| |
is pressed. | loose. |
|
|
| Charge batteries. Refer to Charging the |
| Battery low. | Battery on page 13. |
|
| Rotate RED emergency stop button |
| RED emergency stop button pressed IN. | CLOCKWISE until it pops out. |
|
| Reconnect the battery to the control box. |
| Battery not connected properly to con- | Refer to Charging the Battery on page 13. |
| trol box. |
|
|
| Replace the battery pack. Refer to Charg- |
| The connecting terminals are damaged. | ing the Battery on page 13. |
|
| Refer to Replacing the Boom Actuator on |
| Electric actuator in need of service or | page 22. Contact your Dealer. |
| load is too high. |
|
|
|
|
Unusual noise from actuator. | Actuator is worn or damaged or spindle | Refer to Replacing the Boom Actuator on |
| is bent. | page 22. |
|
|
|
Lift arms will not lower in uppermost | Lift arms require a minimum weight load | Pull down slightly on the lift arms. |
position. | to lower from the uppermost position. |
|
|
|
|
Lift arms will not lower during a power | Shoulder bolt at the junction of the boom | Refer to Replacing the Swivel Bar or |
retraction. | and mast may not be properly installed. | Optional Cradle Attachment on page 25. |
NOTE: If problems are not remedied by the suggested means, please contact your dealer or Invacare.