SECTION 6—TROUBLESHOOTING

SECTION 6—TROUBLESHOOTING

SYMPTOMS

FAULTS

SOLUTION

Patient Lift feels loose.

Mast/Base joint loose.

Refer to Replace the locknut. on page 26.

 

Tie - Rods are loose.

 

 

 

 

Noisy or dry sound from pivots.

Needs lubrication.

Refer to Lubricating the Lift on page 22.

 

 

 

Electric actuator fails to lift when button

Hand-control or actuator connector

 

is pressed.

loose.

 

 

 

Charge batteries. Refer to Charging the

 

Battery low.

Battery on page 13.

 

 

Rotate RED emergency stop button

 

RED emergency stop button pressed IN.

CLOCKWISE until it pops out.

 

 

Reconnect the battery to the control box.

 

Battery not connected properly to con-

Refer to Charging the Battery on page 13.

 

trol box.

 

 

 

Replace the battery pack. Refer to Charg-

 

The connecting terminals are damaged.

ing the Battery on page 13.

 

 

Refer to Replacing the Boom Actuator on

 

Electric actuator in need of service or

page 22. Contact your Dealer.

 

load is too high.

 

 

 

 

Unusual noise from actuator.

Actuator is worn or damaged or spindle

Refer to Replacing the Boom Actuator on

 

is bent.

page 22.

 

 

 

Lift arms will not lower in uppermost

Lift arms require a minimum weight load

Pull down slightly on the lift arms.

position.

to lower from the uppermost position.

 

 

 

 

Lift arms will not lower during a power

Shoulder bolt at the junction of the boom

Refer to Replacing the Swivel Bar or

retraction.

and mast may not be properly installed.

Optional Cradle Attachment on page 25.

NOTE: If problems are not remedied by the suggested means, please contact your dealer or Invacare.