SECTION 8—ACCESSORIES
Part No 1150704 31 Jasmine™ Patient Lift
Calibrating the Jasmine Scale
NOTE:TheJasmineScalewillbepre‐calibratedatthefactorywiththeloadcell.Shoulditbenecessarytore‐calibratethescale,follow
theinstructionsoutlinedbelow.
1. ThepatientandtheslingMUSTberemovedfromthescaletoproperlycalibratetheJasmineScale.Forremovingthe
patientinstructionsrefertoOperationonpage 12.
ForslingdetachmentinstructionsrefertoLiftingthePatient onpage 14.
2. WiththeJasmineScaleon,removethefourscrewsonthebackoftheenclosureandremovethefrontcovertoexpose
thePCboard.
3. PresstheCALbuttonlocatedonthePCboard.TheCALswitchisnotlabeledbutistheonlybuttononthePCboard
locatedinthelowerrightcorner.OncetheCALbuttonispushedʺCAL1ʺwillbeseeninthedisplaywindow.
NOTE:CAL1selectsthecalibrationmodeusing50poundsofcalibratedweight.CAL2selectsthecalibrationmodeforusewith200
poundsofcalibratedweight.PressingtheUNITSkeytogglesbetweenCAL1andCAL2.
4. Whenthedesiredcalibrationmodeisdisplayed,presstheLOCK/UNLOCKkey.
NOTE:ThedisplaywindowwillnowshowUnLd.
5. EnsurethatthereisnoloadonthescaleandpresstheZEROkey.
6. Thedisplaywindowwillshowadashedline(‐‐‐‐‐‐)scrollingacrossandthenthewordLOADwillappearinthedisplay
window.
7. Performone(1)ofthefollowing:
A. FORCAL1OPTION‐support50poundsofcalibratedweightfromthescaleandpresstheZEROkey.
B. FORCAL2OPTION‐support200poundsofcalibratedweightfromthescaleandpresstheZEROkey.
8. Thedisplaywindowwillshowadashedline(‐‐‐‐‐‐)scrollingacrossandthenthewordDONEwillappearinthedisplay
window.
9. RemovetheweightfromthescaleandpresstheLOCK/UNLOCKkey.
NOTE:Theunitwillnowbeinatemporarytestmodeandwillnotlocktheweightdisplay.Thiswillallowweighttobeloadedand
unloadedtocheckthecalibration.
10. TurntheunitoffbypressingtheON/OFFkey.
11. TurntheunitonbypressingtheON/OFFkey.
NOTE:Scalewillnowbeinnormaloperation.
Troubleshooting Display Codes
CALIBRATIONREQUIRED‐Indicatesimproperstoredcalibrationdata,calibrationisnecessary.
OVERCAPACITY‐Indicatesaweightexceedingthecapacityhasbeenloadedonthescale.
SYMPTOM PROBABLE CAUSE SOLUTION
Unit does NOT work properly. Battery failure. Check battery. Replace if necessary.
Battery has been replaced and unit still
does NOT work properly.
Contact Invacare for Service at
1-800-333-6900