Heavy Duty Cleaning Procedure

SECTION 4—SAFETY INSPECTION/TROUBLESHOOTING

Heavy Duty Cleaning Procedure

WARNING

The disinfectant dispensing system has been factory tested and calibrated for use with Invacare disinfectant. The use of any disinfectant with a dilution ratio (½ oz. per gallon) or viscosity level different than that of Invacare disinfectant will result in an improper mixture of disinfectant and water. This may cause your system to be improperly disinfected - or in some instances, not disinfected at all. If using any disinfectant with a dilution ratio or viscosity level different than that of Invacare disinfectant or if switching from one type of disinfectant to another type, the disinfectant flowmeter must be calibrated before performing this procedure. Refer to Calibrating the Disinfectant Flow Meter on page 31.

ALWAYS wear rubber gloves, an apron and face shield protection during disinfecting, cleaning and rinsing procedures. In case of eye or skin contact, follow procedures on disinfectant container.

CAUTION

NEVER use abrasive cleaners like scouring powder or liquid cleaners containing pumice stone. The use of these types of cleaners will make the gel-coat finish of your tub scratched and dull. NEVER use cleaners containing iodine, bromine, Betadine or methylene blue. The use of these types of cleaners will cause the tub to stain. Never use cleaners containing bleach. Use of these types of cleaners will dry out the rubber seals and gaskets and the tub will not function properly.

Before starting the whirlpool pump, the water level must be completely above the whirlpool jets. Otherwise damage to the motor/pump will result.

In addition to disinfecting your tub after every use, you will need to perform this procedure approximately once every 80‐150 baths.

1.Clean the entire surface of the tub to remove visible tissue residue and fluid.

2.Rinse the inside surface of the tub with the shower/massage wand on the spray setting.

3.Rotate the water temperature valve to adjust the water

temperature to approximately 105o F.

NOTE: This will maximize the cleaning solutionʹs ability to clean and disinfect the system.

4.Close the drain.

5.Make sure the whirlpool jets are open.

6.Fill the tub one inch above the highest whirlpool jets.

7.Add the Invacare Heavy Duty Cleaner to the water in the tub. Refer to the chart for the proper amount.

8.Spray pre‐mixed Invacare Heavy Duty Cleaner on the remaining surfaces.

9.Press the whirlpool on/off button to turn on the whirlpool pump and let the pump run for 60 seconds.

10.Turn off the pump and allow the heavy duty cleaner/water solution to remain in the tub for 10 minutes.

11.During this ten‐minute waiting time, take a long handled brush and thoroughly scrub all interior surfaces of the tub.

12.Open the drain to allow the heavy duty/water solution to drain from the tub.

13.Use the shower/massage wand on the pulse setting and direct the spray into both whirlpool jets to back flush until the water discharging from the whirlpool inlet is clear.

TUB

WATER REQUIRED

INVACARE

MODEL

TO COVER JETS

HEAVY DUTY

NUMBER

(GALLONS)

CLEANER

 

 

REQUIRED (OZ)*

 

 

 

2800

10

40

2001

32

128

 

 

 

3700

30

120

3750

30

120

3800

30

120

 

 

 

3600

30

120

3650

48

192

3600XL

48

192

 

 

 

2300

25

100

3300

30

120

4300

35

140

 

 

 

4300HL

35

140

5300

52

208

5300HL

52

208

 

 

 

6300

50

200

6300ADH

50

200

4400SS

48

192

 

 

 

5400SS

50

200

*NOTE: If using non‐Invacare heavy duty cleaner, refer to the instructions supplied with the cleaner to determine the proper amount required.

14.Rinse the tub thoroughly with the shower wand on the spray setting.

15.Check the tub for damage and to ensure all surfaces of the tub are clean and that no visible residue remains. If residue is found, repeat STEPS 13 and 14