SECTION 11—REGULATOR
Perfecto248 Part No 1148070
SECTION 11—REGULATOR

Replacing Regulator

DANGER
Topreventelectricalshock,ALWAYSdisconnectconcentratorfromelectricaloutlet
beforeservicing.
NOTE:Forthisprocedure,refertoFIGURE 11.1onpage49.
1. Unplugunit.
2. Removecabinet.RefertoRemovingCabinetonpage 23.
NOTE:OnconcentratormodelswiththeSensO2unitONLY,removethe1/8inchtubingto
oxygensensorhousing.
3. Removethefourmountingscrewsthatsecureregulatortoproducttankcap.
4. RemoveregulatorfromproducttanklidensuringthattheOringsareremoved.
NOTE:Cleanmountingsurfaceofproducttankcap.EnsurethatthetwoʺOʺringsfortheNEW
regulatorareinstalledontotheports(underside)oftheNEWregulatorbeforeinstallation.
WARNING
Ensureproperinstallationofregulator.Noteflowarrowonregulator.ThisMUSTface
towardtheregulatoroutlettubingorflowwillbeinterruptedcausingsystemshutdown.
5. InstallnewregulatorontoproducttankbyreversingSTEPS3and4.
6. Torquemountingscrewsto5±1inchlbs.
NOTE:OnunitswithSensO2,reinstall1/8inchtubingtooxygensensorhousing.
7. Adjustregulatorpressureifnecessary.RefertoAdjustingRegulatoronpage49.
8. Rununitandcheckforleaks.RefertoLeakTest onpage 76.
9. Reinstallcabinet.RefertoRemovingCabinetonpage 23.