SECTION 18—FLOWMETER
Part No 1148070 63 Perfecto2™
SECTION 18—FLOWMETERReplacing Flowmeter
DANGER
Topreventelectricalshock,ALWAYSdisconnectconcentratorfromelectricaloutlet
beforeservicing.
NOTE:Forthisprocedure,refertoFIGURE 18.1onpage 63.
1. Unplugunit.
2. Removecabinet.RefertoRemovingCabinetonpage 23.
3. NotethelocationofthePVCtubingforaccuratereplacementafternewflowmeteris
installed.
4. Removecontrolpanel.RefertoRemovingControlPanelonpage 53.
5. Removepalnutsthatsecureflowmetertothecontrolpanel.
6. Removeflowmeterfromfrontofcontrolpanel.
7. InstallnewflowmeterreversingSTEPS4‐5.
8. Reinstallcontrolpanel.RefertoRemovingControlPanelonpage 53.
9. Rununitandcheckforleakswhere¼‐inchPVCtubingengagesflowmeterbarbed
fittings.RefertoLeakTest onpage 76.
10. Reinstallcabinet.RefertoRemovingCabinetonpage 23.
FIGURE 18.1 Replacing Flowmeter
Palnuts
Control Panel
Flowmeter Barbed
Fittings (Back of
Flowmeter)
Flowmeter