SECTION 21—PURITY INDICATORS
Part No 1148070 71 Perfecto2
SECTION 21—PURITY INDICATORS

SensO2 Alarm Indicators

DANGER
Topreventelectricalshock,ALWAYSdisconnectconcentratorfromelectricaloutlet
beforeservicing.
NOTE:Forthisprocedure,refertoFIGURE 21.1onpage 72.

Operation

ThePerfecto2withSensO2Concentratorisequippedwithanoxygenpurityindicator.
Thisfeaturemonitorsthepurityleveloftheoxygengeneratedbytheconcentrator.If
purityfallsbelowalarmthresholds,indicatorlightsonthecontrolpanelwillilluminate.
NOTE:Concentratormaybeusedduringtheinitialstartwarmuptime(approximately30
minutes)whilewaitingfortheO2puritytoreachmaximum.
Whentheunitisturnedon,theGREENlightwillcomeon(SYSTEMOK/O2greaterthan
85%).Afterfiveminutes,theoxygensensorwillbeoperatingnormallyandwillcontrol
theindicatorlightsdependingonoxygenconcentrationvalues.Theexplanationofthe
indicatorlightfunctionsareasfollows:

Indicator Lights

NOTE:Indicatorlightsarevisibleonlywhenlit.
GREENlight(O2)‐NormalOperation.
YELLOWlight()‐CallsupplierIMMEDIATELY.Youmaycontinuetousethe
concentratorunlessinstructedotherwisebyyoursupplier.Becertainthatbackupoxygen
isnearby.
REDlight()‐Total unitshutdown.SwitchIMMEDIATELYtobackupoxygensupply.
CallsupplierIMMEDIATELY.
GREENlight‐withYELLOWlightflashing‐CallsupplierIMMEDIATELY.Oxygen
sensormalfunctioning;youmaycontinuetousetheconcentrator.Indicators‐Ifyourunit
doesnotfeaturetheO2Sensor

LX Models Indicator Light Explanation

REDlight()‐TotalUnitShutDown.SwitchIMMEDIATELYtoabackupoxygen
supply.CallSupplierIMMEDIATELY.
GREENlight()‐On/Off.Systemokay.
I/O