SECTION 24—ALARM TEST
Perfecto2™84 Part No 1148070
P.E. Valve Coil TestNOTE:Forthisprocedure,refertoFIGURE 24.4.
NOTE:P.E . Va lve Coil‐AlarmsequenceoccurswhentheP.E . Va lv eCoil,connectionorP.C .
BoardCircuitryhasfailed.
1. Withtheunitrunningandflowsetat5L/min.,removeoneYELLOWwirefromthe
P.E . valve coil.TheP.E . valvecoilalarmshouldactivatewithin10seconds.
2. Reinstallcabinet.RefertoRemovingCabinetonpage 23.
NOTE:Ifanyalarmfailstoperformtospecification,contactInvacareTechnicalService.
NOTE:Ifproblemsariseduringtest,contactInvacareTechnicalServiceforfurtherassistance.
FIGURE 24.4 P.E. Valve Coil Test
UsethechartinPreventiveMaintenanceonpage24torecorddateandnumberofhours
whenpreventivemaintenancewasperformedontheconcentratororanyrepairsmade.
Yellow Wires
P.E. Valve Coil